Reading an Accurate Blank
Why blanking matters in UV/Vis spectroscopy
What does a blank do?
During a UV/Vis measurement, the spectrophotometer records the light intensity that passes through the sample. The instrument compares this transmitted light to the initial light intensity and calculates absorbance using the Beer-Lambert relationship.
In practical terms, the blank tells the instrument what “zero” should look like. The blank spectrum is stored as the background and subtracted from later sample readings.
That means your sample result depends not only on the sample itself, but also on the quality of the blank.
Why a bad blank causes bad data
A compromised blank can introduce false peaks, negative absorbance, distorted spectra, or incorrect concentration values.
Common causes include contaminated water or buffer, residue on the measurement window or mirror, air bubbles from pipetting, foam-forming detergents in buffers, aging or chemically changed buffer solutions, and residual alcohol after cleaning.
A good blank should produce a flat, horizontal baseline when the blank solution is measured again as a sample.
Best-practice blanking workflow
-
- Start with fresh deionized water Apply 1–1.5 µL of fresh deionized water to the measurement window and run a blank. Avoid water from open containers, even if labeled RNase-free.
- Check the water as a sample Remove the blank with a clean, dry wipe. Apply fresh water again and measure it as a sample. The result should be a flat baseline.
- Measure your buffer as a sample Apply the buffer and check whether it absorbs in the wavelength range of interest. If the buffer produces peaks near your sample’s target wavelength, it may not be suitable.
- Blank with buffer only after verification Once the buffer is confirmed not to interfere, record it as the blank. Then reapply fresh buffer and measure it as a sample to confirm a flat baseline.
- Use fresh tips and reverse pipetting Use a new pipette tip at every step. Reverse pipetting helps reduce bubbles and foam, especially with small volumes, proteins, nucleic acids, detergents, or foaming buffers.
Preventing Measurement Interference and Contamination
Buffers and reagents to watch carefully
Some buffer components absorb strongly in the UV range and can interfere with protein or nucleic acid measurements. These include RIPA, Tween 20, NDSB, and Triton X-100.
If these substances create absorbance peaks in the wavelength range of interest, they may compromise your measurement.
Cleaning the measurement surfaces
Clean the measurement window and mirror with a moistened wipe first, then dry thoroughly. For stubborn contamination, use a 70/30 alcohol/water solution, followed by a water-moistened wipe and then a dry wipe.
Do not leave alcohol residue behind, because alcohol can absorb in the higher UV range.